Low Temperature Centre Sticks
25 Jun 2013
Yes
-  

 

 

 

The following are the types of Gas Handling Low temperature centre sticks that we have on offer.

No
100mm Gas Handling Centre Stick
 

If you have any questions or want to use one of these sticks in your experiment, please contact the pressure and furnace group.

On cryostat centre sticks the capillary has a temperature sensor and heater to ensure that the sample gas does not freeze, preventing blockages. The centre stick is terminated at the sample end with a standard 34mm Conflat flange facilitating attachment of the sample container with a standard copper gasket. The Centre Stick capillary is connected via a 1/8", (3.175mm) diameter tube and, Swagelok fittings to the gas handling rig.

 

Gas Handling Centre stick Description Working parameters Number of
Gas handling centre sticks 100mm Gas handling centre stick for use in 100mm cryostat or top loading CCR. Vacuum to 200bar. 15
Gas handling centre sticks 50mm Gas handling centre stick for use in 50mm cryostat or top loading CCR. Vacuum to 200bar. 4
Volumetric centre sticks 100mm Gas handling centre stick for use in 100mm cryostat or top loading CCR when conducting volumetric experiments. Vacuum to 200bar. 3
Cryo furnace gas handling centre sticks 70mm Gas handling centre stick for use in 70mm cryofurnace. Vacuum to 200bar. 2
Mari gas handling centre stick  Gas handling centre stick for use in 100mm cryostat or top loading CCR with Mari horizontal configuration. Vacuum to 200bar. 1
Contact: