Quantum Design Physical Property Measurement System-9
01 Oct 2020
Yes
-  

 

 

 

No

​​​The PPMS System-9 in the Materials Characterisation Laboratory​​​

 

Overview​

The PPMS System-9 is an open architecture, variable temperature and field system designed to perform a variety of scripted and automated magnetic and physical property m​easurements.​​


Common Research Applications

  • Magnetic properties: magnetic DC and AC susceptibility, magnetometry

  • Thermal properties: heat capacity, thermal transport, cryo-FMR

  • Electric properties: DC and AC resistance, electro transport​

​​​

Specifications

  • Capable of 24/7 continuous operation

  • Measurement temperature range of 1.8 to 400 K with measurement sweep rate of 0.01 to 10 K/min

  • Continuous Low-Temperature Control (CLTC) for precise and uninterrupted operation below 4.2 K

  • Applied field strength of ± 9 Tesla

  • Field uniformity of ± 0.01 % over 5.5 cm

​​

Measurement Options

Heat Capacity

The Heat Capacity option for the PPMS will run fully automated relaxation heat capacity measurements under high-vacuum.

AC Susceptibility

The AC Measurement System (ACMS) option can perform both AC susceptibility and DC magnetisation measurements without changing between hardware. A single automated measurement system is capable of both.

  • AC frequencies from 10 Hz to 10 kHz and field amplitudes from 0.05 Oe to 15 Oe

  • AC susceptibility measurement sensitivity of 1 x 10-8 emu at 10 kHz

  • DC magnetisation measurement sensitivity of 5 x 10-6 emu

  • ACMS brochure

AC Transport

The AC Transport (ACT) option support four different automated electrical transport modes: AC resistivity, five-wire Hall effect, I-V curves and critical current measurements.

  • Two samples can be loaded simultaneously

  • Current range of 10 μA to 2 A

  • Sensitivity of 1 nV at 1 kHz

  • Operational frequency range of 1 Hz to 1 kHz

DC Resistivity

The DC Resistivity option adds resistance bridge capabilities to the instrument providing four independent channels for van der Pauw and four-wire resistance measurements.

  • Up to three samples can be loaded and measured simultaneously

  • Current range of 5 nA to 5 mA

  • 20 nV sensitivity

  • DC Resistance brochure

Ferromagnetic Resonance

The Cryo-FMR probe enables measurement of the Ferromagnetic Resonance (FMR) response of a material at variable temperatures and sample orientations parallel or perpendicular to the DC field.

Manufacturer's website​​